Sodium Formate Chromatographic Conditions and System Configuration info@pulisichem.cn
After optimization, a typical set of determination conditions is provided below. Analysts may fine-tune these based on the performance of the actual instrument and chromatographic column:
Sodium Formate Mobile Phase: 20 mM potassium dihydrogen phosphate buffer (pH 3.0) : methanol = 95:5 (v/v). The buffer must be filtered through a 0.45 µm membrane filter and degassed. The pH significantly affects the peak shape and retention time of sodium formate.
Chromatographic Column: Reversed-phase C18 column, 250 mm × 4.6 mm, 5 µm. Ensure the column is in good condition with acceptable efficiency.
Flow Rate: 1.0 mL/min (constant flow).
Sodium Formate Column Temperature: 30°C, maintained within the 25–40°C range for analytical stability.
Detection Wavelength: 210 nm. Sodium formate exhibits end absorption near this wavelength; ensure the detector’s deuterium lamp has sufficient energy.
Injection Volume: 10 µL. Auto-samplers are recommended; if using manual injection, pay close attention to operational consistency.
We Shandong Pulisi Chemical Co., Ltd. specialize in the production and distribution of premium grade sodium formate. Featuring excellent solubility and stable chemical properties, this product serves as ideal raw materials for construction admixtures, leather industry, printing and dyeing industry as well as feed additives. We maintain sufficient inventory and offer flexible delivery solutions to meet diverse purchasing demands from overseas markets.